pH Measurement: Importance, Applications, and Techniques

pH Measurement is used in a wide variety of applications: agriculture, wastewater treatment, industrial processes, environmental monitoring, and research and development.

pH is a measure of the Acidity or Alkalinity of a Solution. The pH value states the relative quantity of hydrogen ions (H+) contained in a solution. The greater the concentration of H+ the more acidic the solution and the lower the pH. In this relationship, pH is defined as the negative logarithm of hydrogen activity.

pH sticks in a lab

pH Measurement

A pH below 7 is acidic, with lower numbers indicating higher acidity. A pH above 7 is alkaline, with higher numbers indicating greater alkalinity. pH is an important factor in many processes. The pH scale ranges from 0 to 14. A pH of 7 is neutral, meaning it has an equal concentration of hydrogen ions and hydroxide ions (OH-).

For example, it can affect the activity of enzymes, the availability of nutrients for plants, and the solubility of many substances. It can also affect the taste of food and drinks and can be used to measure the acidity or alkalinity of water and soil. In industry, pH is used to monitor and control many processes, including the production of fertilizers, drugs, and food products. A precise pH measurement is essential for many industrial processes and is often monitored and controlled with automated systems.

Value of pH

The pH value of a solution is the negative log of its hydrogen ion activity (α), which is the product of hydrogen ion concentration [H+] and the activity coefficient of hydrogen (gH+) at that concentration.

pH = –log α = –log γH+[H+]

In pure water and in dilute solutions, the H+ activity can be considered the same as the H+ concentration.

pH = –log γH+[H+] = –log [H+]

The pH of a solution measures the degree of acidity or alkalinity relative to the ionization of water. Pure water dissociates to yield 10-7 M of [H+] and [OH–] at 25°C; thus, the pH of water is 7, the point of neutrality.

pHwater = –log [H+] = –log 10-7 = 7

Most pH readings range from 0 to 14. Solutions with a higher [H+] than water (pH less than 7) are acidic; solutions with a lower [H+] than water (pH greater than 7) are basic or alkaline.

Definition of pH

Measuring pH involves comparing the potential of solutions with unknown [H+] to a known reference potential. pH meters convert the voltage ratio between a reference half-cell and a sensing half-cell to pH values. Today, most electrodes are combination electrodes with both the reference and sensing half-cells in the same body. Reference Half-Cells contain a conductor (usually silver with a silver chloride coating) immersed in a solution with known [H+]. The potential between this internal conductor and the known solution is constant, providing a stable reference potential.

Sensing Half-Cells (measuring half-cells) are made of a nonconducting glass (or epoxy) tube sealed to a conductive glass membrane. Like the reference half-cell, the sensing half-cell also contains a conductor immersed in a buffered electrolyte solution, ensuring constant voltages on the inner surface of the glass membrane and the sensing conductor.

When the pH electrode is immersed in the solution to be measured, a potential is established on the surface of the sensing glass membrane. If the unknown solution is neutral, the sum of fixed voltages on the inner surface of the glass membrane and on the sensing conductor approximately balances the voltage on the outer surface of the glass membrane and the reference half-cell. This results in a total potential difference of 0 mV and a pH value of 7.

In acidic or alkaline solutions, the voltage on the outer membrane surface changes proportionally to changes in [H+]. The pH meter detects the change in potential and determines [H+] of the unknown by the Nernst equation where:

  • E = total potential difference (measured in mV)
  • T = temperature in Kelvin
  • n = number of electrons
  • E° = reference potential
  • F = Faraday’s constant
  • R = gas constant
  • [H+] = hydrogen ion concentration

pH Temperature Compensation

Typical pH Electrode Response as a Function of TemperatureTypical pH Electrode Response as a Function of Temperature

Typical pH Electrode Response as a Function of Temperature

The pH of any solution is a function of its temperature. The voltage output from the electrode changes linearly in relationship to changes in pH, and the temperature of the solution determines the slope of the graph. One pH unit corresponds to 59.16 mV at 25°C, the standard voltage and temperature to which all calibrations are referenced. The electrode voltage decreases to 54.20 mV/pH unit at 0.0°C and increases to 74.04 mV/pH unit at 100.0°C.

Since pH values are temperature dependent, pH applications require some form of temperature compensation to ensure standardized pH values. Meters and controllers with automatic temperature compensation (ATC) receive a continuous signal from a temperature element and automatically correct the pH value based on the temperature of the solution. Manual temperature compensation requires the user to enter the temperature. ATC is considered to be more practical for most pH applications.

Sources:

Schmuckler, J. S. PH Concept. J. Chem. Educ. 1982, 59 (4), 305.https://doi.org/10.1021/ed059p305.2.

Kolb, D. The PH Concept. J. Chem. Educ. 1979, 56 (1), 49. https://doi.org/10.1021/ed056p49.

Request A Quote

Customer Information
Shipping Address
Billing Address

Will you need a lift gate? *

(For deliveries requiring a LTL truck and the customer does not have a dock door)

Would you like to add shipping amount to quote? *

Are you tax exempt? *

(If you are tax exempt, please email your tax exemption form to [email protected])

Upload PDF: