Autoclaving Laboratory Glassware

Autoclaving Laboratory Glassware

Autoclaves are widely used to sterilize instruments, glassware and plasticware, solutions and media and to decontaminate biological wastes. Because of the physical hazards (e.g., heat, steam and pressure) associated with autoclaving, extra care ust be taken to ensure their safe use. Each autoclave has unique characteristics and operating requirements. Review and understand the owner’s manual before using any autoclave for the first time and as needed thereafter. Also follow all of your institution’s safety regulations for autoclaving.

Autoclaving PYREX® Glassware
Corning recommends the following safety practices when autoclaving laboratory glassware:

  • Never autoclave items containing corrosives (e.g., acids, bases, phenol), solvents or volatiles (e.g., ethanol, methanol, chloroform) or radioactive materials.
  • Use only borosilicate glassware, such as PYREX glass,which can better withstand the stresses of high autoclave temperatures and pressures.
  • Load the autoclave properly as per the manufacturer’s recommendations.
  • Individual glassware vessels should be placed within a heat resistant plastic or metal tray on a shelf or rack and never placed directly on the autoclave bottom or floor. There should always be space (>5 cm) left between glass vessels to give them room to expand during heating and to prevent them hitting other vessels when entering or leaving the autoclave.
  • Add 1/4 to 1/2 inch of water to the tray so the glassware will heat more evenly.
  • Check any plastic caps, tubing or other items to ensure they can be safely autoclaved with the glassware.
  • Fill glassware only half full with liquids to be sterilized.  Take into account the volume of liquid to be autoclaved.  A 2 liter flask containing 1 liter of liquid takes much longer to sterilize than four 500 mL flasks each containing 250 mL of liquid.
  • Do a test run when autoclaving larger volumes of liquid to make sure the liquid has reached 121ºC. Suspend an autoclave thermometer (with a thin wire) in the middle of the liquid filled container to record the highest temperature reached. Then check its reading after the autoclave run is finished, if the temperature is too low then the run time will need to be increased or the volume will need to be reduced.
  • Large heavy glassware, such as 3L or larger spinner flasks, 5L or larger bottles and carboys,should have a small amount of distilled water placed inside them to help generate steam when they are being sterilized dry in an autoclave.
  • To prevent bottles from shattering during pressurization, the caps of vessels with liquids must be fully loosened before loading. Always use the slow exhaust or liquid cool cycle when autoclaving liquids to prevent liquid boil over which can result in loss of content.
  • Never autoclave a sealed vessel containing liquids as this may result in an explosion of super-heated liquid and steam during the cycle or when the vessel is opened.
  • Wear heat-resistant gloves when opening the autoclave door after a cycle. At a minimum, when removing items from an autoclave, a rubber apron, rubber sleeve protectors and heat-resistant gloves should be worn.
  • Wait 5 minutes before removing dry glassware from the autoclave, and wait 10 to 20 minutes for glassware containing liquids. The larger the volume of liquid the longer you should wait. Liquids removed too soon may boil and overflow the vessel, burning operator.
  • DO NOT tighten caps on vessels immediately after autoclaving as the vacuum resulting from the cooling of the steam in the vessel can cause the cap or vessel to shatter.  This is especially important on larger vessels. Caps with liners may have their liner sucked into the vessel if the cap is not lose enough on the vessel opening.
  • Do not put hot glassware, especially large bottles, from an autoclave (or other heat source) on to a cold bench top.  The stress that is induced will often cause the vessel bottom
    to crack or fall off releasing its hot contents.
  • For non-liquid loads, let the glassware cool for 15 minutes before touching it with ungloved hands.
  • For liquid loads, let liquids stand for a full hour before touching with ungloved hands. Be sure others in the area know a heat hazard is present.


Autoclaving PYREXPLUS® Glassware
PYREXPLUS glassware has a protective vinyl coating that requires some additional precautions for autoclaving. PYREXPLUS glassware can be successfully sterilized repeatedly using liquid or dry cycle sterilization which involves no vacuum or low vacuum (less than 5 inches mercury) cycles.

  • Sterilization time should not exceed 15 minutes at 121°C (250°F).
  • It is better to use a slow cool cycle for venting rather than fast venting to reduce the chance that steam or air pockets will form between the glass and the coating.
  • Drying time should not exceed 15 minutes at 110°C (230°F). The actual cavity temperature of the autoclave
    should be checked to be sure the autoclave temperature does not exceed the recommended sterilization and drying temperature.
  • Vessels should not be allowed to touch each other during autoclaving.
  • Should the coating appear clouded due to dissolved moisture, simply let dry overnight at room temperature or briefly heat to 110°C (230°F).

     

Information courtesy of CORNING

Related Categories
livechat