STM-E Sterilmatic® Digital Sterilizer Autoclaves
Manufacturer: Market Forge Industries
Description
To ensure compatibility with your facility's power infrastructure, the Sterilmatic series is available in several electrical configurations. Please verify your voltage and phase requirements after ordering.
| Model | kW | Hz | Phase | Voltage | Amperage |
|---|---|---|---|---|---|
STM-E |
9.3 |
60 |
1 Phase |
208V |
45A |
STM-E |
9.3 |
60 |
3 Phase |
208V |
26A |
STM-EL |
12.4 |
60 |
1 Phase |
240V |
52A |
STM-EL |
12.4 |
60 |
3 Phase |
240V |
30A |
Features:
- The spacious chamber allows for the sterilization of large flasks, beakers, and surgical kits that typically require a much more expensive, floor-standing conventional sterilizer
- Equipped with a timer, steam gauge, low water cut-off system, a signal light to indicate cycle status, and an ASME-certified safety valve to prevent over-pressurization (CSA Approved)
- Features a thermostatic steam trap and removable pan supports, making it easy to clean the interior and maintain the unit's long-term reliability
- The digital interface provides more consistent results than traditional analog dials, ensuring your protocols meet strict regulatory standards for sterilization
- The warranty is 90 Days Labor, 1 Year Parts
Operational & Maintenance Insights:
- Power Requirements: Due to the high wattage (up to 12.4 kW), these units require a dedicated circuit. Ensure your facility's wiring and circuit breakers are rated for the specific amperage.
- Water Quality: To prevent mineral buildup on the heating elements and internal baffle, always use distilled or deionized water. Tap water can lead to "scaling," which reduces heating efficiency and may trigger the low-water cut-off prematurely.
- Steam Baffle: Ensure the flat perforated steam baffle is correctly positioned before starting a cycle. This ensures even steam distribution and prevents "cold spots" in the chamber.
- Door Seal: Regularly inspect the door gasket for cracks or debris. A clean seal is essential for reaching and maintaining the 30 lb. steam pressure required for a successful cycle.
Please Note: The options and accessories are for use with Market Forge STERILMATIC Steam Sterilizers (Autoclaves). The accessories can be used with STM-ED (95-6300), STM-E (95-2678), and STM-EL (95-3441) Models - unless otherwise noted.
| SKU | Material | Max Temperature | Temperature Range | Warranty | External Dimensions |
|---|---|---|---|---|---|
| 95-2678 | Stainless Steel | 260°F | 250°F (121°C) | 1 Year | 32" x 18" x 27" |
| 95-3441 | Stainless Steel | 260°F | 230 -250°F | 1 Year | 32" x 18" x 27" |
|
|
Sterilmatic® Steam Sterilizer Product Manual | Download |
|
|
Sterilmatic® Steam Sterilizer Autoclave Parts List | Download |
|
|
Sterilmatic® Steam Sterilizer Specification Sheet | Download |
-
Read MoreGood Laboratory Technique demands clean Glassware, because the most carefully executed piece of work may give an erroneous result if dirty glassware is used.
-
Read MoreAutoclaves are widely used to sterilize instruments, glassware and plasticware, solutions and media, and to decontaminate biological wastes.
Starting at $19,771.82
Select a product below and add it to your cart to request a quote.