High-Performance Filter Holders
Manufacturer: GVS
Description
Features:
- The exterior locking ring and inlet cap design allow for quick assembly by hand-tightening without the risk of tearing or twisting the membrane filter
- Volume Specific Optimization:
- 13 mm Swinney: Optimized for small volumes (1–5 mL) dispensed via syringe
- 25 mm: Designed for filtering up to 50 mL of sample
- 47 mm: Capable of filtering up to 1 Liter, depending on sample viscosity
- The 47 mm model conforms to EPA Method 1311 (TCLP)* for hazardous waste compliance, essential for evaluating metal mobility (Regulatory Compliance)
- In-line holders feature dual support screens, allowing for Bi-Directional Flow and preventing membrane rupture during back-pressure events
- Several syringe holders can be attached together for multi-stage serial filtration processes
Applications:
- Analysis of Biological and Organic Fluids
- Biofuel analysis
- Ophthalmics
- Gas chromatography samples
- Lubricants
Maintenance & Operation:
- IMPORTANT! Autoclaving Note: To ensure proper steam penetration and prevent thermal stress, reassemble the unit loosely before placing it in the autoclave
- Sealing: 3 O-rings to prevent leaks across all membrane types; ensure O-rings are seated properly before hand-tightening
- Cleaning: Disassemble and clean with non-abrasive detergents. Rinse thoroughly with deionized water and air dry before reassembly
Conforms with EPA Method 1311 for Toxicity Characteristic Leaching Procedure 40 CFR, Part 261, 1991 Hazardous Waste Compliance Guide
| SKU | Connector Type | Material | Max Pressure | Max Temperature | Filtration Area |
|---|---|---|---|---|---|
| 1214250 | Female Luer Lock Inlet, Male Luer Slip Outlet | Polypropylene | 2.9 bar (42 psi) | 80°C (176°F) | 3.5 cm2 |
| 1220950 | Female Luer Lock Inlet, Male Luer Slip Outlet | Celcon (Acetal Copolymer) | 2.8 bar (40 psi) | 80°C (176°F) | 0.8 cm2 |
| 1214260 | 1/4 in NPTM Female Luer Slip, 1/4 in NPTM Female Luer Slip | Polypropylene | 4.9 bar (71 psi) | 80°C (176°F) | 13.5 cm2 |
| 1262579 | 1/4 in NPTM Female Luer Slip, 1/4 in NPTM Female Luer Slip | Polypropylene | 4.9 bar (71 psi) | 80°C (176°F) | 13.5 cm2 |
|
|
GVS Membrane Filter Holder Specification Sheet | Download |
| SKU | Autoclavable | Includes |
|---|---|---|
| 1214250 | 20 minutes at 121°C (250°F) and 15 psi | Silicone O-Rings |
| 1220950 | 15 minutes at 121°C (250°F) and 15 psi | PTFE O-Rings |
| 1214260 | 20 minutes at 121°C (250°F) and 15 psi | Silicone O-Rings |
| 1262579 | 20 minutes at 121°C (250°F) and 15 psi | Silicone O-Rings |
Starting at $195.00