Certificate of Calibration for Serialized Class A Glass

Accurate, reliable scientific work relies on precision in measurement wherever it is applied. This includes equipment like balances and scales but also manual tools like graduated cylinders or burettes. Volumetric glassware calibration helps ensure the measuring accuracy of laboratory glass pieces wherever they are used prior to being sold. There are also recommendations available for the recalibration of older pieces to ensure they are still measuring accurately after a certain period of time, based on the type of glass.

Glassware

Below you will find an example of a certificate created for a calibrated piece of serialized Class A glassware manufactured by Glassco. This type of certificate describes the calibrated piece of glass, the test method, the item batch, and the testing results.

CALIBRATION CERTIFICATE

Glassco Calibration Laboratory ISO 9001:2000 Certified

USP-STANDARD

    CERTIFICATE NO.: GLASSCO/10/07-02/508

    CRF. DATE: 01/07/2010

    CAL. REPORT NO.: 10/07-02

    CUSTOMER: UN1-3996

    TEST METHOD: GRAVIMETRIC METHOD

    ITEM CALIBRATED: MEASURING CYLINDER H/BASE: 100 ml

    CATALOGUE NO.: 139.220.04A

    BATCH NO. & SERIAL NO. OF MEASURING CYLINDER: 05.10/508

    ITEM CALIBRATED ON: 01/07/2010

    CALIBRATION RESULT:

    The product was found to contain the following volume at 20ºC.

    VOLUME CONTAINED: 100.000 ml

    >TOLERANCE: +/-0.500 ml

    CORRECTION: +0.000 ml

    STANDARD DEVIATION: 0.001 ml

    Expanded uncertainty (at 95%C.L.k=2) is 0.001 ml

    The product has been calibrated with the following devices:

    Standard weights: 100g

    Using Balance Number: ABAL-03, 420g/0.001g (SARTORIUS)

    Thermometer NO.: DM-05; -50ºC TO 150ºC

    All the above devices have valid calibrations traceable to NPL India; traceable to international standards.  The above product has been examined for compliance to ASTM E-1272 Class-A.  It has been found to satisfy the specified requirements. The above item has been calibrated at 25.1ºC at RH 39%.

    The above calibration result is valid at the time of calibration under lab conditions. This report shall not be reproduced, except in full, unless written permission from GLASSCO LABORATORY EQUIPMENTS.

    OPERATOR'S NAME: Monica
    PREPARED BY: Sangeet
    APPROVED AND ISSUED BY: Manager QA Mr. Gagan Bagga

Request A Quote

Customer Information
Shipping Address
Billing Address

Will you need a lift gate? *

(For deliveries requiring a LTL truck and the customer does not have a dock door)

Would you like to add shipping amount to quote? *

Are you tax exempt? *

(If you are tax exempt, please email your tax exemption form to [email protected])

Upload PDF: