Certificate of Calibration for Serialized Class A Glass - Example
CALIBRATION CERTIFICATE(Glassco Calibration Laboratory ISO 9001:2000 Certified)(USP-STANDARD)CERTIFICATE NO.: GLASSCO/10/07-02/508 CRF. DATE: 01/07/2010 CAL. REPORT NO.: 10/07-02 CUSTOMER: UN1-3996 TEST METHOD: GRAVIMETRIC METHOD ITEM CALIBRATED: MEASURING CYLINDER H/BASE: 100 ml CATALOGUE NO.: 139.220.04A BATCH NO. & SERIAL NO. OF MEASURING CYLINDER: 05.10/508 ITEM CALIBRATED ON: 01/07/2010 CALIBRATION RESULT: The product was found to contain the following volume at 20ºC. VOLUME CONTAINED: 100.000 ml TOLERANCE: +/-0.500 ml CORRECTION: +0.000 ml STANDARD DEVIATION: 0.001 ml Expanded uncertainty (at 95%C.L.k=2) is 0.001 ml The product has been clibrated with the following devices: Standard weights: 100g Using Balance Number: ABAL-03, 420g/0.001g (SARTORIUS) Thermometer NO.: DM-05; -50ºC TO 150ºC All the above devices have valid calibrations traceable to NPL India; traceable to international standards. The above product has been examined for compliance to ASTM E-1272 Class-A. It has been found to satisfy the specified requirements. The above item has been calibrated at 25.1ºC at RH 39%. The above calibration result is valid at the time of calibration under lab conditions. This report shall not be reporduced, except in full, unless written permission from GLASSCO LABORATORY EQUIPMENTS. OPERATOR'S NAME: MONICA PREPARED BY: SANGEET APPROVED AND ISSUED BY: Manager QA Mr. Gagan Bagga Related Categories
|